Spectral domain optical coherence tomography imaging with an integrated optics spectrometer
| Authors |
|
|---|---|
| Publication date | 2011 |
| Journal | Optics Letters |
| Volume | Issue number | 36 | 7 |
| Pages (from-to) | 1293-1295 |
| Organisations |
|
| Abstract |
We designed and fabricated an arrayed-waveguide grating (AWG) in silicon oxynitride as a spectrometer for spectral domain optical coherence tomography (SD-OCT). The AWG has a footprint of only 3.0 cm x 2.5 cm, operates at a center wavelength of 1300 nm, and has 78 nm free spectral range. OCT measurements are performed that demonstrate imaging up to a maximum depth of 1 mm with an axial resolution of 19 mu m, both in agreement with the AWG design parameters. Using the AWG spectrometer combined with a fiber-based SD-OCT system, we demonstrate cross-sectional OCT imaging of a multilayered scattering phantom. (C) 2011 Optical Society of America
|
| Document type | Article |
| Language | English |
| Published at |
https://doi.org/10.1364/OL.36.001293
(Final published version)
|
| Downloads |
364677.pdf
(Final published version)
|
| Permalink to this page | |