Operating the GridPix detector in dark matter search experiments

Authors
Publication date 2013
Journal Nuclear Instruments & Methods in Physics Research Section A - Accelerators Spectrometers Detectors and Associated Equipment
Volume | Issue number 718
Pages (from-to) 446-449
Organisations
  • Faculty of Science (FNWI) - Institute of Physics (IoP) - Institute for High Energy Physics (IHEF)
Abstract
The DARWIN (dark matter WIMP search with noble liquids) design study aims to use liquid argon and liquid xenon targets to look for nuclear recoils due to weakly interacting massive particles (WIMPs). To measure the recoil energy in dual-phase noble gas time projection chambers the combination of scintillation and ionisation detection is used to discriminate nuclear from electron recoils. Current experiments use an array of photomultiplier tubes to detect the primary scintillation and the ionisation electrons via secondary scintillation in the gas phase. Within the research framework for DARWIN, one candidate for an alternative direct charge readout is GridPix, a micro-pattern gaseous detector composed of a Micromegas-like amplification grid over the Timepix 65k pixel readout chip. It can achieve a single-electron detection efficiency of up to 98% and has thus great potential to identify the ionisation electrons in dark matter search experiments. The main challenges for this application are low outgassing, thermal robustness, and operation in pure (thus quencher-free) noble gas. To investigate its applicability we operated a GridPix detector in an argon cryostat. We proved GridPix's performance in pure argon. The gas amplification was confirmed in pure argon in the broad temperature range from 300 K down to 87 K. Additionally, we discuss results of thermal tests of GridPix devices at liquid xenon temperature in a dry nitrogen atmosphere.
Document type Article
Language English
Published at https://doi.org/10.1016/j.nima.2012.11.054
Permalink to this page
Back