The heat-pipe resembling action of boiling bubbles in endovenous laser ablation
| Authors |
|
|---|---|
| Publication date | 2010 |
| Journal | Lasers in Medical Science |
| Volume | Issue number | 25 | 6 |
| Pages (from-to) | 907-909 |
| Organisations |
|
| Abstract |
Endovenous laser ablation (EVLA) produces boiling bubbles emerging from pores within the hot fiber tip and traveling over a distal length of about 20 mm before condensing. This evaporation-condensation mechanism makes the vein act like a heat pipe, where very efficient heat transport maintains a constant temperature, the saturation temperature of 100A degrees C, over the volume where these non-condensing bubbles exist. During EVLA the above-mentioned observations indicate that a venous cylindrical volume with a length of about 20 mm is kept at 100A degrees C. Pullback velocities of a few mm/s then cause at least the upper part of the treated vein wall to remain close to 100A degrees C for a time sufficient to cause irreversible injury. In conclusion, we propose that the mechanism of action of boiling bubbles during EVLA is an efficient heat-pipe resembling way of heating of the vein wall
|
| Document type | Article |
| Published at |
https://doi.org/10.1007/s10103-010-0780-2
(Final published version)
|
| Downloads |
341664.pdf
(Final published version)
|
| Permalink to this page | |