The heat-pipe resembling action of boiling bubbles in endovenous laser ablation

Open Access
Authors
  • C.W.M. van der Geld
  • R.R. van den Bos
  • P.W.M. van Ruijven
  • T. Nijsten
  • H.A.M. Neumann
  • M.J.C. van Gemert
Publication date 2010
Journal Lasers in Medical Science
Volume | Issue number 25 | 6
Pages (from-to) 907-909
Organisations
  • Faculty of Medicine (AMC-UvA)
Abstract
Endovenous laser ablation (EVLA) produces boiling bubbles emerging from pores within the hot fiber tip and traveling over a distal length of about 20 mm before condensing. This evaporation-condensation mechanism makes the vein act like a heat pipe, where very efficient heat transport maintains a constant temperature, the saturation temperature of 100A degrees C, over the volume where these non-condensing bubbles exist. During EVLA the above-mentioned observations indicate that a venous cylindrical volume with a length of about 20 mm is kept at 100A degrees C. Pullback velocities of a few mm/s then cause at least the upper part of the treated vein wall to remain close to 100A degrees C for a time sufficient to cause irreversible injury. In conclusion, we propose that the mechanism of action of boiling bubbles during EVLA is an efficient heat-pipe resembling way of heating of the vein wall
Document type Article
Published at
https://doi.org/10.1007/s10103-010-0780-2 (Final published version)
Downloads
341664.pdf (Final published version)
Permalink to this page