Comparison of five cone beam computed tomography systems for the detection of vertical root fractures

Authors
Publication date 2010
Journal The Journal of Endodontics
Volume | Issue number 36 | 1
Pages (from-to) 126-129
Organisations
  • Faculty of Dentistry (ACTA)
Abstract
Introduction
This study compared the accuracy of cone beam computed tomography (CBCT) scans made by five different systems in detecting vertical root fractures (VRFs). It also assessed the influence of the presence of root canal filling (RCF), CBCT slice orientation selection, and the type of tooth (premolar/molar) on detection accuracy.
Methods
Eighty endodontically prepared teeth were divided into four groups and placed in dry mandibles. The teeth in groups Fr-F and Fr-NF were artificially fractured; those in groups control-F and control-NF were not. Groups Fr-F and control-F were root filled. CBCT scans were made using five different commercial CBCT systems. Two observers evaluated images in axial, coronal, and sagittal reconstruction planes.
Results
There was a significant difference in detection accuracy among the five systems (p = 0.00001). The presence of RCF did not influence sensitivity (p = 0.16), but it reduced specificity (p = 0.003). Axial slices were significantly more accurate than sagittal and coronal slices (p = 0.0001) in detecting VRF in all systems. Significantly more VRFs were detected among molars than premolars (p = 0.0001).
Conclusions
RCF presence reduced specificity in all systems (p = 0.003) but did not influence accuracy (p = 0.79) except in one system (p = 0.012). Axial slices were the most accurate in detecting VRFs (p = 0.0001).

Document type Article
Language English
Published at https://doi.org/10.1016/j.joen.2009.09.013
Permalink to this page
Back