Optimizing computer-assisted orbital reconstruction From digital design to clinical care
| Authors |
|
|---|---|
| Supervisors | |
| Co-supervisors |
|
| Award date | 18 November 2026 |
| ISBN |
|
| Number of pages | 165 |
| Organisations |
|
| Abstract |
Orbital fractures can cause enophthalmos, hypoglobus and diplopia. Accurate reconstruction of the complex orbital anatomy is essential to restore function and appearance. Computer-assisted surgery can be conceptualized as a modular toolbox spanning four phases: advanced diagnostics, virtual surgical planning, surgical transfer and evaluation. This thesis aimed to optimize and clinically translate a personalized workflow for orbital reconstruction.
Part I focuses on patient-specific implants (PSIs). Design strategies for three- and four-wall defects are described, including multiple interlocking implants. A virtual comparison of PSI designs for identical patients showed that both surgeon and engineer shape the final design, with a median difference of 22% between surgeons. This underlines the need for evidence-based design protocols. Part II focuses on surgical navigation. A virtual splint registration workflow was clinically validated in ten patients. Pre-existing fixation screws enabled measurement of the target registration error at the surgical site: 1.5 mm. A real-time navigation workflow was developed with industrial partners, in which a tracked instrument is attached to the implant. Six visualization methods were evaluated to optimize its feedback. In a cadaver study, real-time navigation achieved accuracy comparable to conventional navigation. It reduced reconstruction time from 9.5 to 4.2 minutes and halved the number of attention shifts, measured with eye-tracking. This thesis shows that PSI design is shaped not only by the patient’s anatomy and pathology, but also by surgeon and engineer. It also shows that clinical translation requires dedicated methods to measure what is often assumed. Rigorous validation and patient-relevant outcomes should guide further development. |
| Document type | PhD thesis |
| Language | English |
| Downloads |
Thesis (complete)
(Embargo up to 18 November 2028)
Chapter 7: Preclinical validation of real-time navigation
(Embargo up to 18 November 2028)
|
| Supplementary materials | |
| Permalink to this page | |