Validation of photoplethysmography using a mobile phone application for the assessment of heart rate variability in the context of heart rate variability-biofeedback

Open Access
Authors
  • W. van Dijk
  • A.C. Huizink
  • M. Oosterman
  • I.L.J. Lemmers-Jansen
Publication date 09-2023
Journal Psychosomatic Medicine
Volume | Issue number 85 | 7
Pages (from-to) 568-576
Number of pages 9
Organisations
  • Faculty of Social and Behavioural Sciences (FMG) - Research Institute of Child Development and Education (RICDE)
Abstract

OBJECTIVE: Heart rate variability-biofeedback (HRV-BF) is an effective intervention to reduce stress and anxiety and requires accurate measures of real-time HRV. HRV can be measured through photoplethysmography (PPG) using the camera of a mobile phone. No studies have directly compared HRV-BF supported through PPG against classical electrocardiogram (ECG). The current study aimed to validate PPG HRV measurements during HRV-BF against ECG. METHODS: Fifty-seven healthy participants (70% women) with a mean (standard deviation) age of 26.70 (9.86) years received HRV-BF in the laboratory. Participants filled out questionnaires and performed five times a 5-minute diaphragmatic breathing exercise at different paces (range, ~6.5 to ~4.5 breaths/min). Four HRV indices obtained through PPG, using the Happitech software development kit, and ECG, using the validated NeXus apparatus, were calculated and compared: RMSSD, pNN50, LFpower, and HFpower. Resonance frequency (i.e., optimal breathing pace) was also compared between methods. RESULTS: All intraclass correlation coefficient values of the five different breathing paces were "near perfect" (>0.90) for all HRV indices: lnRMSSD, lnpNN50, lnLFpower, and lnHFpower. All Bland-Altman analyses (with just three incidental exceptions) showed good interchangeability of PPG- and ECG-derived HRV indices. No systematic evidence for proportional bias was found for any of the HRV indices. In addition, correspondence in resonance frequency detection was good with 76.6% agreement between PPG and ECG. CONCLUSIONS: PPG is a potentially reliable and valid method for the assessment of HRV. PPG is a promising replacement of ECG assessment to measure resonance frequency during HRV-BF.

Document type Article
Note With supplementary file.
Language English
Published at https://doi.org/10.1097/PSY.0000000000001236
Other links https://www.scopus.com/pages/publications/85170195454
Downloads
Supplementary materials
Permalink to this page
Back