Cavitation measurement during sonic and ultrasonic activated irrigation

Authors
  • R. Macedo
  • B. Verhaagen
  • D. Fernandez Rivas
  • M. Versluis
Publication date 2014
Journal The Journal of Endodontics
Volume | Issue number 40 | 4
Pages (from-to) 580-583
Organisations
  • Faculty of Dentistry (ACTA)
Abstract
Introduction
The aims of this study were to quantify and to visualize the possible occurrence of transient cavitation (bubble formation and implosion) during sonic and ultrasonic (UAI) activated irrigation.

Methods
The amount of cavitation generated around several endodontic instruments was measured by sonochemiluminescence dosimetry inside 4 root canal models of human dimensions and varying complexity. Furthermore, the spatial distribution of the sonochemiluminescence in the root canal was visualized with long-exposure photography.

Results
Instrument oscillation frequency, ultrasonic power, and file taper influenced the occurrence and amount of cavitation. In UAI, cavitation was distributed between the file and the wall extending beyond the file and inside lateral canals/isthmuses. In sonic activated irrigation, no cavitation was detected.

Conclusions
Cavitation was shown to occur in UAI at clinically relevant ultrasonic power settings in both straight and curved canals but not around sonically oscillating instruments, driven at their highest frequency.
Document type Article
Language English
Published at https://doi.org/10.1016/j.joen.2013.09.018