Multidetector CT imaging of mechanical prosthetic heart valves: quantification of artifacts with a pulsatile in-vitro model

Open Access
Authors
  • P. Symersky
  • R.P.J. Budde
  • P. Westers
  • B.A.J.M. de Mol
  • M. Prokop
Publication date 2011
Journal European Radiology
Volume | Issue number 21 | 10
Pages (from-to) 2103-2110
Organisations
  • Faculty of Medicine (AMC-UvA)
Abstract
Multidetector computed tomography (MDCT) can detect the cause of prosthetic heart valve (PHV) dysfunction but is hampered by valve-induced artifacts. We quantified artifacts of four PHV using a pulsatile in-vitro model and assessed the relation to leaflet motion and valve design. A Medtronic Hall tilting disc (MH), and Carbomedics (CM), St Jude (SJM), and ON-X bileaflet valves underwent CT in an in-vitro model using retrospective gating with a 64 detector CT system in stationary and pulsatile conditions. Artifacts and radiopaque component volumes were quantified with thresholds based on surrounding structures and valvular components. Hypodense artifacts volumes (mm³) were 1,029 ± 147, 535 ± 53, 371 ± 16, and 366 ± 18 for the SJM, MH, CM and ON-X valves (p < 0.001 except for the latter two valves p = 0.43). Hyperdense artifact volumes were 3,546 ± 141, 2,387 ± 103, 2,003 ± 102, and 3,033 ± 31 for the SJM, MH, CM and ON-X valve, respectively (all differences p < 0.001). Leaflet motion affected hypodense (F = 41.5, p < 0.001) and hyperdense artifacts (F = 53.7, p < 0.001). Closed and moving leaflets were associated with the least and the most artifacts respectively (p < 0.001, both artifact types). Both valve design and leaflet motion affect PHV-induced artifacts. Best imaging results may be expected for the CM valve during phases in which the leaflets are closed
Document type Article
Language English
Published at
https://doi.org/10.1007/s00330-011-2146-y (Final published version)
Downloads
365299.pdf (Final published version)
Permalink to this page
Back